Tibial Fracture Fixation Team Logo.jpg
FA info icon.svg Angle down icon.svg Medical equipment data
Parent 3D Printed Adult Male Humeral Bone Models

Caption underneath

Prepare 3D Print Files[edit | edit source]

Have these instructions displayed on a mobile device to verify that all the steps are properly performed when preparing the print files on a computer.

OOjs UI icon notice-destructive.svg

Please pay attention to and follow all the instructions closely to ensure the bone models are printed properly and display the required visual, tactile, and acoustic fidelity for orthopedic surgical simulation training.

OOjs UI icon notice-destructive.svg

Do not change the filament material and do not deviate from these published print settings because they have been pre-tested to ensure the fidelity and printability of these surgical simulators.

1
Download STL Files

Download the 3D files (.STL) in the table below to input the print settings for the specific 3D printer to create the print files (.GCODE).

Adult Male Humeral Bone Model 3D Files Model # Download STL File Revision Date
1 Click on this link and click on "Model_1_-_Male_-18-Oct-2022_version.stl" to download the file. October 18, 2022
2 Click on this link and click on "Model_2_-_Male_-18-Oct-2022_version.STL" to download the file. October 18, 2022
2
Install 3D Slicer Program

You will need to download, install, and use the latest version of:

  • Ultimaker Cura, the world’s most popular 3D printing software with millions of users, which already has profiles for many commercially available 3D printers, allows you to create a custom profile for other 3D printers, and is open-source and free to download,[1] or
  • Cura LulzBot Edition if you are using any of the open-source, open filament LulzBot 3D printers.
OOjs UI icon notice-destructive.svg

If your 3D printer cannot print g-code files created using Ultimaker Cura or Cura Lulzbot Edition, you cannot print the Adult Humeral Bone Models. Without Ultimaker Cura or Cura Lulzbot Edition software, you will not be able to input the custom print setting options, the bone models will not print properly and the simulators will not display the required visual, tactile, and acoustic fidelity for orthopedic surgical simulation training.

3
Open the 3D Slicer Program.png
Open the 3D Slicer Program

Open the 3D Slicer Program.

4
Add 3D Printer.png
Add 3D Printer

Click on "Add printer."

5
Add A Non-Networked Printer.png
Add A Non-Networked Printer

Click on "Add a non-networked printer."

6
Select 3D Printer.png
Select 3D Printer
  • Scroll down to find and select the name of your 3D printer.
  • If you don't see your 3D printer, you can select "Custom" and fill in the "Printer" and "Printhead" settings for your 3D printer.
  • Click on "Add."
7
Select Filament Material.png
Select Filament Material
  • For the filament material, select the brand name of "PLA" that you will be using.
  • If your PLA filament brand is not an available option, select "Generic PLA" as the filament material.

The bone models must be 3D printed using white polylactic acid (PLA) filament because PLA filament has a hardness value that is very similar to human cortical bone.[2][3][4][5]

OOjs UI icon notice-destructive.svg

Do not substitute PLA filament with acrylonitrile butadiene styrene (ABS), engineering co-polyester (CPE+), or glycol-modified polyethylene terephthalate (PETG) filament because our December 26, 2021 internal testing found that 3D printed bone models made of Ultimaker ABS filament (Shore Hardness 76D), and Ultimaker CPE+ (Shore Hardness 77D) were unable to tolerate powered or manual drilling of a self-drilling 4.5 mm diameter Schanz Screw during simulation training, and our research has found that Ultimaker PETG has a Shore Hardness of 76D.[6][7][8][9][10]

OOjs UI icon notice-destructive.svg

If you have previously modified any values from the 3D slicer program's default settings for PLA filament, please change these values back to the default settings by clicking on the circular arrow to the left of each modified setting.

For optimal print results and to minimize the risk of mechanical failure of the bone models during orthopedic surgical simulation training, always use brand new white PLA filament fresh out of the sealed package.[11]

8
Open Ultimaker Cura - Male Model 1.png
Verify Nozzle Diameter
  • Select the nozzle diameter that matches the nozzle installed on the 3D printer.
  • The default nozzle diameter is 0.4 mm.
9
Open Slicer Program.png
Import STL File
  • Click on "File."
  • Click on "Open File."
  • Select the filename of the .STL file you want to import.
  • Click on "Open."
10
Select Model - Male Model 1.png
Select Model

Click on the imported 3D model on the build plate.

11
Rotate Model - Male Model 1.png
Rotate Model If Required
  • If required, rotate the model so the base with the semi-engraved model number, gender symbol, and drilling direction arrow is lying flat on the build plate.
  • Model 1 usually requires rotation by 180 degrees in the X-axis (shown in red).
12
Lay Flat - Male Model 1.png
Apply Lay Flat Feature

Click on "Lay Flat" to ensure the model is lying flat on the build plate.

13
Lay Flat - Male Model 1.png
Rescale Models If Required
  • If the 3D printer has a build volume Z height of at least 170 mm, all 3D Printed Adult Humeral Bone Models can be printed at 100% and no re-scaling is required.
  • If the 3D printer has a maximum build volume Z height between ~150 mm to 170 mm, all 3D Printed Adult Humeral Bone Models must be uniformly and identically re-scaled at less than 100% and at the highest scale factor which permits printing on the specific 3D printer.
  • The Ultimaker Cura 5.0.0 program shows the 3D model with grey stripes to indicate that the model does not fit inside the 3D printer's build volume and needs to be re-scaled.

For uniform scaling at the highest scale factor for both Models 1 and 2, we recommend scaling Model 2 before Model 1 because Model 2 is slightly larger than Model 1.

14
Re-Scale Model - Male Model 1.png
Select Scale Feature
  • If your 3D printer has a build volume Z height between ~150 to 170 mm, click on the 3D model on the build plate.
  • Click on "Scale."
  • Uniformly scale the model to the highest scale factor which permits printing on the specific 3D printer.
  • With the "Uniform Scaling" box checked, input a percent value less than 100% (i.e., 95%) in the X row and press "Enter" on your keyboard to uniformly re-scale the model.
  • Click on "Lay Flat" to ensure the model is lying flat on the build plate.
  • Check if the re-scaled model fits in the 3D printer's build volume. In the Ultimaker Cura 5.0.0 program, the 3D model's colour will change from grey stripes to solid yellow to indicate that the re-scaled model now fits in the 3D printer's build volume.
  • Repeat the prior 3 steps above with different percent values (i.e., 94%, or 96%, or 97%, etc.,) until you identify the highest percent value that permits printing of the re-scaled model in the 3D printer.
  • Record the scale factor (i.e., 95%) and apply the identical scale factor for the other models.
  • To deselect the model, click anywhere in the slicer screen outside of the model.
15
Set Recommended Print Settings - Male Model 1.png
Set Recommended Print Settings
  • Profile: To minimize print times, use the lowest quality or fastest speed (layer height of 0.15 mm to 0.3 mm or less) recommended for the specific 3D printer
  • Support: none (unchecked)
  • Click on "Custom."

To avoid print quality issues, the layer height should be 0.15 mm to 0.3 mm or less.

OOjs UI icon notice-destructive.svg

Do not select the "Support" option because support material will cover up the semi-engraved model number, gender symbol, and drilling direction arrows on the base of each model designed to assist with model identification and proper orientation.

16
Set Wall Thickness

Walls -> Wall Thickness: 5.5 mm (this changes the Wall Line Count)

The wall thickness value is set to the average far cortex thickness for the distal end of the humeral shaft for male adults ages 21-45.[12]

17
Adjust Top Layer Setting - Male Model 1.png
Adjust Top Layer Setting
  • Top Layers: 0
  • Bottom Layers: default value (not 0)

Since Models 1 and 2 are simulated fracture ends, then the Top Layers should be "0" for Models 1 and 2 to allow the tibial bone's interior anatomy to be exposed at each fracture end.

OOjs UI icon notice-destructive.svg

Do not change the Bottom Layers to "0" because this will make the base open at the bottom.

18
Select Infill Pattern - Male Model 1.png
Select Infill
  • Infill -> Infill Density: 15%
  • Infill -> Infill Pattern: Tri-Hexagon

The infill density is set to simulate human cancellous bone porosity.[13] The infill pattern is selected to simulate human cancellous bone microstructure.[14]

OOjs UI icon notice-destructive.svg

Do not increase the print speed beyond the slicer program's default generic (unmodified) settings for PLA. If the print speed is too high, it will cause filament drooping.[15] If you have previously modified the print speed values from the slicer program's default settings for PLA, please change the print speed values back to the default settings by clicking on the circular arrow to the left of each modified setting.

19
Reduce Top-Bottom Speed.png
Reduce Top/Bottom Speed
  • Speed -> Top/Bottom Speed: 15.0 mm/s

The Top/Bottom Speed is reduced to help ensure adhesion of the base of the model to the build plate during printing and minimize print failures.

20
Select Build Plate Adhesion v2.0.png
Select Build Plate Adhesion
  • Build Plate Adhesion -> Build Plate Adhesion Type: Skirt (default) or Brim (follow the build plate manufacturer's recommendations for using a brim to ensure adhesion of PLA to the build plate).
OOjs UI icon notice-destructive.svg

Do not select the "Raft" option because a raft will cover up the semi-engraved model number, gender symbol, and drilling direction arrows on the base of each model designed to assist with model identification and proper orientation.

21
Slice Model
  • Click on "Slice."
  • Click on "Save to Disk."
  • Save the G-CODE file to avoid errors when inputting the customized print settings and eliminate print file preparation time when re-printing in the future.
  • Record the "Printing Time" and "Filament Weight" values to calculate a price quote.

Please proceed to the next step by clicking on this link or by clicking on the "Next" button in the Menu box in the upper right section of this page.

Acknowledgements[edit | edit source]

This work is funded by a grant from the Intuitive Foundation. Any research, findings, conclusions, or recommendations expressed in this work are those of the author(s), and not of the Intuitive Foundation.

References[edit | edit source]

  1. https://ultimaker.com/software/ultimaker-cura
  2. www.prusa3d.com/file/370474/technical-data-sheet.pdf
  3. Ultimaker. Ultimaker PLA Technical Data Sheet [Internet]. Ultimaker Support. [cited 2021 July 29]. Available from: https://support.ultimaker.com/hc/en-us/articles/360011962720-UltimakerPLA-TDS.
  4. Vian, Wei Dai and Denton, Nancy L., "Hardness Comparison of Polymer Specimens Produced with Different Processes" (2018). ASEE IL-IN Section Conference. 3. https://docs.lib.purdue.edu/aseeil-insectionconference/2018/tech/3
  5. Society For Biomaterials 30th Annual Meeting Transactions, page 332. Femoral Cortical Wall Thickness And Hardness Evaluation. K. Calvert, L.A. Kirkpatrick, D.M. Blakemore, T.S. Johnson. Zimmer, Inc., Warsaw, IN.
  6. https://support.ultimaker.com/hc/article_attachments/4800377696540/Ultimaker-ABS-TDS-v5.00.pdf
  7. https://support.ultimaker.com/hc/en-us/article_attachments/360026416800/ultimaker-PETG-SDS-v1.00.pdf
  8. https://support.ultimaker.com/hc/en-us/article_attachments/360010204279/SDS_CPE__v3.004-en.pdf
  9. https://support.ultimaker.com/hc/article_attachments/4800535847708/Ultimaker-CPE-PLUS-TDS-v5.00.pdf
  10. https://support.ultimaker.com/hc/en-us/article_attachments/4777237744540/Ultimaker-PETG-TDS-v1.00.pdf
  11. https://support.ultimaker.com/hc/en-us/articles/360012101319-How-to-store-material
  12. Meema HE, Meema S. Measurable roentgenologic changes in some peripheral bones in senile osteoporosis.J Am Geriat Soc 1963;11:1170-82.
  13. Forrest AM, Johnson AE, inventors; Pacific Research Laboratories, Inc., assignee. Artificial bones and methods of making same. United States patent 8,210,852 B2. Date issued 2012 Jul 3.
  14. National Institutes of Health Osteoporosis and Related Bone Diseases National Resource Center. What is Bone? [Internet]. Bethesda (MD): The National Institutes of Health (NIH); 2018. [Cited 2021 Aug 17]. Available from: https://www.bones.nih.gov/health-info/bone/bone-health/what-is-bone.
  15. https://all3dp.com/2/3d-printing-overhang-how-to-master-overhangs-exceeding-45/
Cookies help us deliver our services. By using our services, you agree to our use of cookies.