No edit summary
mNo edit summary
Line 1: Line 1:
{{MOST}}
{{Lit}}
[[Category:MOST literature reviews]]
*[http://rsi.aip.org/resource/1/rsinak/v60/i9/p3051_s1 A computercontrolled turbidity apparatus]
*[http://rsi.aip.org/resource/1/rsinak/v60/i9/p3051_s1 A computercontrolled turbidity apparatus]
**I_t = I_0e^{-(α+τ)l}
**I_t = I_0e^{-(α+τ)l}

Revision as of 15:43, 5 June 2013

  • A computercontrolled turbidity apparatus
    • I_t = I_0e^{-(α+τ)l}
    • We're looking for τ.
    • Aparatus for measuring turbidity as function of temperature; includes oven to place sample in. Claims 0.001/mm accuracy and 0.5% precision.
    • They measure transmitted light and assume no absorbtion; use a laser of polarized light with beam splitter.
    • Use chopper to measure ac instead of dc, claim order of magnitude better long-term stability from it.
    • Much improvement from simultaneous reference and measurement beam.
  • Comparison Type Nephelometer
    • Measure transmitted and scattered light with an integrating sphere each, and direct intensity, all three at once.
    • Polarized, old style (1952), not really useful.