How to measure the accuracy/reproducibility of a syringe pump using a digital mass scale.
 +
Accuracy/Reproducibility Protocol:
 +
AFTER CALIBRATION
 +
1. Use a liquid with a known density (water) to test this.
 +
2. Put some water in the syringe.
 +
3. Push out 5mL of water onto a scale. Record the mass in grams.
 +
4. Repeat step 3 nine more times.
 +
5. Calculate the volume of water that came out each trial. (Volume = Mass/Density)
 +
6. Compare each trial the volume of each trial to 5mL. Reproducibility = Actual/Theoretical x 100%
 
[[Category:MOST methods]]
 
[[Category:MOST methods]]

Revision as of 20:34, 7 December 2013


Sunhusky.png By Michigan Tech's Open Sustainability Technology Lab.

Wanted: Students to make a distributed future with solar-powered open-source 3-D printing.
Currently looking for PhD or MSC student interested in solar energy policy- apply now!
Contact Dr. Joshua Pearce - Apply here

MOST: Projects & Publications, Methods, Lit. reviews, People, Sponsors, News
Updates: Twitter, Instagram, YouTube

OSL.jpg

How to measure the accuracy/reproducibility of a syringe pump using a digital mass scale. Accuracy/Reproducibility Protocol: AFTER CALIBRATION 1. Use a liquid with a known density (water) to test this. 2. Put some water in the syringe. 3. Push out 5mL of water onto a scale. Record the mass in grams. 4. Repeat step 3 nine more times. 5. Calculate the volume of water that came out each trial. (Volume = Mass/Density) 6. Compare each trial the volume of each trial to 5mL. Reproducibility = Actual/Theoretical x 100%