===Accuracy/Reproducibility Protocol:=== | ===Accuracy/Reproducibility Protocol:=== | ||
====Using a iM01 My Weigh Scale==== | ====Using a iM01 My Weigh Hich Precision Scale==== | ||
AFTER CALIBRATION | AFTER CALIBRATION | ||
#Use a liquid with a known density (water) to test this. | #Use a liquid with a known density (water) to test this. |
Revision as of 16:02, 13 December 2013
Accuracy/Reproducibility Protocol:
Using a iM01 My Weigh Hich Precision Scale
AFTER CALIBRATION
- Use a liquid with a known density (water) to test this.
- Put some water in the syringe.
- Push out 5mL of water onto a scale. Record the mass in grams.
- Repeat step 3 nine more times.
- Calculate the volume of water that came out each trial. (Volume = Mass/Density)
- Compare each trial the volume of each trial to 5mL. Reproducibility = Actual/Theoretical x 100%