| Bandwidth | Channels | Real time
sampling rate |
Capture rate | Memory depth | |
| SDS1104X-U | 100 MHz | 4 | 1 GSa/s | 4 00,000 wfm/s | 14 Mpts |
Please perform the following steps to clean the instrument and probe regularly according to its operating conditions.
3. Equipment specifications:
Calibration should be done after 180 days past the Date of Calibration" on the certificate with the unit and after each calibration interval.
The self-calibration program can quickly make the oscilloscope reach the best working state to get the most precise measurement values. You can perform self-calibration at any time especially when the change of the environment temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.
Do the following steps to do self calibration:


When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy or error. The probe compensation procedures are as follows.

1. Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation.
2. Connect to Power Supply
The power requirements of the oscilloscope are 100-240 V, 50/60/440 Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source.
3. Power-on Inspection
When the oscilloscope is energized, press the power key at the lower-left corner of the front panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-tests and you can hear the sound of relay switching. After the self-test is finished, the welcome screen is displayed.
4. Connect the Probe
SIGLENT provides passive probes for the oscilloscopes.
5. Run Control
Auto Setup: press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display.
Run/Stop: press the button to set the acquisition state to Run or Stop.
“Index of /upload_file/user/SDS1000X-E.” https://int.siglent.com/upload_file/user/SDS1000X-E/ (accessed Apr. 03, 2023).
| License | CC-BY-SA-4.0 |
|---|---|
| Organizations | FAST |
| Cite as | Motakabbir (2023–2025). "SDS1104X-E 100Mhz Digital Oscilloscope: FAST". Appropedia. Retrieved September 28, 2026. |