SDS1104X-E 100Mhz Digital Oscilloscope
Bandwidth Channels Real time

sampling rate

Capture rate Memory depth
SDS1104X-U 100 MHz 4 1 GSa/s 4 00,000 wfm/s 14 Mpts

Safety Training Requirements

Safety Instructions

Working Environment

Temperature

Ventilation

Care

Cleaning

Please perform the following steps to clean the instrument and probe regularly according to its operating conditions.

SDS1104X-E 100Mhz Digital Oscilloscope

  1. Location: FAST laboratory, Room TEB 06, Thomson Engineering Building, Western University, London, Ontario, Canada
  2. Vendor: Siglent Technologies
Contact details
  • Name: SIGLENT Technologies North America, Inc
  • Address: 6557 Cochran Rd Solon, Ohio 44139
  • Phone: Tel: 440-398-5800; Toll Free:877-515-5551; Fax: 440-399-1211
  • Email: support@siglentna.com
  • Website: siglentna.com

3. Equipment specifications:

Calibration & Tolerances

Calibration and Due date for calibration

Calibration should be done after 180 days past the Date of Calibration" on the certificate with the unit and after each calibration interval.

Self Calibration

The self-calibration program can quickly make the oscilloscope reach the best working state to get the most precise measurement values. You can perform self-calibration at any time especially when the change of the environment temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.

Do the following steps to do self calibration:

  1. Disconnect all the input channels.
  2. Press the Utility button on the front panel and then press the Do Self Cal softkey.
  3. Press the Single button on the front panel to perform the self calibration program. During the calibration, most of the keys are disabled.
  4. When the self calibration program is finished, it will pop-out the message ―press Run/Stop key to exit‖. Press the Run/Stop button on the front panel to exit the calibration interface.

Function Inspection

  1. Press the Default button on the front panel to restore the instrument to its default configuration.
  2. Connect the ground alligator clip of the probe to the ―Ground Terminal‖ under the probe compensation signal output terminal.
    Compensation Signal
  3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the Compensation Signal Output Terminal‖ of the probe.
  4. Press the Auto Setup softkey.
  5. Observe the waveform on the display. In normal condition, the display should be a square waveform as shown in the figure below:
    Function inspection

Probe Compensation

When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy or error. The probe compensation procedures are as follows.

  1. Set the switch to 10X on the probe.
  2. Perform steps 1, 2, 3 and 4 of ―Function Inspection‖ in the previous section.
  3. Check the waveforms displayed and compare them with the following:
    Compensation
  4. Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole on the probe until the waveform displayed is as the ―Perfectly compensated‖ in the figure above.

Operation & Procedure

1. Adjust the Supporting Legs

Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation.

2. Connect to Power Supply

The power requirements of the oscilloscope are 100-240 V, 50/60/440 Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source.

3. Power-on Inspection

When the oscilloscope is energized, press the power key at the lower-left corner of the front panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-tests and you can hear the sound of relay switching. After the self-test is finished, the welcome screen is displayed.

4. Connect the Probe

SIGLENT provides passive probes for the oscilloscopes.

5. Run Control

Auto Setup: press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display.

Run/Stop: press the button to set the acquisition state to Run or Stop.

References

“Index of /upload_file/user/SDS1000X-E.” https://int.siglent.com/upload_file/user/SDS1000X-E/ (accessed Apr. 03, 2023).

    Page data
    Keywords safety, laboratory
    License CC-BY-SA-4.0
    Organizations FAST
    Language English ()
    Related 0 subpages, 1 pages link here
    Views 61 page views (analytics)
    Created April 3, 2023 by Md. Motakabbir Rahman
    Last edit June 12, 2025 by Felipe Schenone