The photoconductivity measurements are performed using the QAS Solar Simulator. This type of measurement provides a current reading for a range of voltages.

Protocol

  • For Test Platform the wires correspond to the following: White is the purple wire and therefore the brass coloured spring loaded lead. Red is the red alligator clip which should be connected to the metal loop attached to the base of the sample platform. This lead allows a voltage / current to be applied to the whole sample. The top left black alligator clip corresponds to the green lead. The bottom left black alligator clip corresponds to the black lead. These two black leads attach to the probes which should make contact with the top surface of the sample.
  • NOTE: The leads are in pairs white and red, black and green and as long as these pairing in maintained you can switch the set-up.
  • NOTE: The two probe ends need to be as close together as possible without touching and they need to be on a contact on the cell. Having the probes just on the cell will not produce quality results and if the cells are not close together, one may get erratic results.
  • For Meter cable, the other end of the cable has the same colour leads and therefore as long as one has both the white and red as positive leads and the green and black as negative leads the system will work (or vice versa).

To obtain the I-V curve:

  • Press the toggle button - config - Source V - arrow one to the right so that the sensing mode is blinking - enter - chose 2 or 4 wire
  • Press the range to increase/decrease the maximum compliance value. One can move along the digits with the edit horizontal arrows and to increase/decrease the digits value press the edit vertical arrows.
  • Press Edit button and the Iscr or Vscr should start blinking. Use the edit horizontal arrows to move to the desired digit and then press the edit vertical arrows to increase/decrease the applied current/voltage.
  • At a certain value the recorded value should exceed the compliance, this would be the maximum value, the other end of the range would be were the recorded value does not change any further.

Dark current measurements: