The Integrating Sphere: IntegratIR Mid-Infrared Integrating Sphere module is designed to sit in the sample compartment of the FTIR spectrometer, and connect to it's internal detector port. It is used to measure diffuse reflectance and diffuse transmittance of solids, but can be adapted through the use of a sample plate with a ZnSe window to accept powders as well.
Make sure FTIR spectrophotomer that you are installing the integrating sphere in is the one it was intended to work with. Operation with a different FTIR may require changes that can only be completed by the manufacturer (PIKE). The base plate makes sure that the integrating sphere is properly lined up so as not to interfere with the incoming beam.
You are now ready to collect data using the integrating sphere.
| Parameter | Setting |
|---|---|
| Wavelength Range | 5000 to 600 cm-1 |
| Aperture | Open |
| Resolution | 4 or 8 cm-1 |
| Number of scans | 64 |
| Velocity | 1.8 - 2.5 cm/sec |
Taylor Method of Sample Measurement to Avoid Substitution Error - METHOD OF CHOICE!!
Another method - we have not found it to work. Place Gold reference on window for background spectra and alignment. After use replace cover on gold surface to avoid contamination Sample Prep & Measurements
Parameters that have a significant effect on the likelihood of obtaining the same spectrum with every measurement:
Special Considerations