Line 87: Line 87:


   *'''FiO2:''' 1.0 (100%)
   *'''FiO2:''' 1.0 (100%)
===Regulations===
* Contact FDA for how to go about getting device cleared.  They have a policy for ventilators here: https://www.fda.gov/media/136318/download
** If your design is a modification of a design already cleared by FDA I think that would make it easier to get approval for.  It might be worth it to also give them a call just to be sure?
** I think what they would give you (or what you should aim for) is an Emergency Use Authorization: https://www.fda.gov/emergency-preparedness-and-response/mcm-legal-regulatory-and-policy-framework/emergency-use-authorization 
** A ventilator would fall under other medical devices: https://www.fda.gov/emergency-preparedness-and-response/mcm-legal-regulatory-and-policy-framework/emergency-use-authorization#covidothermeddev  and they have an email for you to contact CDRH-NonDiagnosticEUA-Templates@fda.hhs.gov
===Safety Precautions===
===Safety Precautions===
* Confirm that the wiring in the ventilator is correctly set-up (see wiring diagram)
* Confirm that the wiring in the ventilator is correctly set-up (see wiring diagram)

Revision as of 13:54, 2 April 2020

Over the rest of the semester MOST and well as the MTU Open Source Hardware Enterprise will be documenting, building and testing open source ventilators. If you want to help please do.

Background

Follow literature review protocols from the group here:[1]

How Many are Needed?

See Dave Denkenbergers' spreadsheet

https://www.mdedge.com/rheumatology/article/219591/coronavirus-updates/fda-allow-alternative-respiratory-devices-treat

Existing Designs

MTU Designs

Testing Procedures

MTU Testing Procedure

  • Select on of the four modes using the user interface on the ventilator:
 *Continuous mechanical ventilation (CMV): used when a patient is not breathing on their own
 *Inverse ratio ventilation (IRV): used when a patients inhale lasts longer than their exhales(i.e 2:1 instead of 1:2)
 *Pressure support ventilation (PSV): when a patient has trouble completing a breath thus resulting in negative pressure during a breathing circuit
 *Assist Control (AC): used when a patient needs support for every breath
  • Confirm that the peak inspiratory pressure (PIP), respiratory rate (RR), positive end expiratory pressure (PEEP), inspiratroy:expiratory ration (i:e), and tidal volume (TV) are accurate by using the Michigan Instruments software package that accompanies the lung.
  • Test the alarms by unplugging the machine, wiring, and appropriate tubing. The accessories that should be tested are:
 *Low Pressure Alarm: Decrease the PIP value using the ventilators user interface to an abnormal value.
 
 *High Pressure Alarm: Increase the PIP value using the ventilators user interface to an abnormal value.
 
 *Oxygen Concentration Alarm: Increase/decrease the percentage of oxygen per breathe to above/below the safety range (19.5-23.5%) by mechanically increasing/decreasing the amount of O2 entering the ventilator and using an O2 sensor. 
 
 *Battery Backup Alarm: Unplug the ventilator's primary battery from the voltage source.

 *Wire Disconnection Alarm: Unplug a wire from the circuit board in the ventilator while it is running. 
 
 *Oxygen Tube Disconnection Alarm: Unplug the oxygen intake tube from the ventilator while the machine is on.
  • Ideal Values that Should be Targeted for a Healthy Patient:

https://www.nursingcenter.com/getattachment/Clinical-Resources/nursing-pocket-cards/Mechanical-Ventilation-Settings-and-Basic-Modes/Mechanical-Ventilation-Settings-and-Basic-Modes-Tip-Card_January-2019.pdf.aspx

 *PIP: 10-14 cm H2O
 
 *RR: 8-12 breathes per minute (BPM)
 *PEEP: 3-10 cm H2O
 *I:E: 1:2
 *TV: 6-8 mL/kg
 *O2 Concentration: 19.5-23.5%
 *FiO2: 1.0 (100%)

Regulations

Safety Precautions

  • Confirm that the wiring in the ventilator is correctly set-up (see wiring diagram)
  • Make sure that the ventilator is properly connected to the artificial lung with the proper adapter size (see picture below)
  • Confirm that the artificial lung is set-up correctly BEFORE turning the machine on (see manual below, pg.#)